Chem On Singapore: GESTER Masks Clogging Tester GT-RA13

Products

GESTER Masks Clogging Tester GT-RA13

  • LOGIN FOR MORE INFO

The sample is surrounded by a known concentration of dolomite dust in air. Following the exposure, the breathing resistance and the filter penetration of the sample particle filtering half mask are measured.

Product Introduction

Application:

 

The sample is surrounded by a known concentration of dolomite dust in air. Following the exposure, the breathing resistance and the filter penetration of the sample particle filtering half mask are measured.

 

Standards:

EN 149 section 8.10


Share

Model

 GT-RA13

Dust

 DRB 4/5 dolomite

Dust Distributor

 Particle size range: 0.1μm~10μm

 Mass Flow Range: 40mg/h---400mg/h

Breathing Machine

 Displacement:2.0L/stroke

 Frequency: 10times/min

Exhaled Air Temperature

 (37±2)°C

Exhaled Air Humidity

 95% R.H minimum

Continuous Flow Through the dust chamber

 60 m3/ h, linear velocity: 4cm/s

Concentration of the Dust

 (400±100)mg / m3

Test Chamber

 Inner Dimension: 650 mm×650 mm×700 mm

 Airflow: 60 m3/ hlinear velocity: 4cm/s

 Air Temperature: (23±2)°C

 Air Humidity: (45±15)

Respiratory Resistance Measurement Range

 02000Pa, accuracy: 0.1Pa

Power Supply

 220V, 50Hz, 1KW

External Dimension

 800mm×600mm×1650mm