The Large Direct-Pure EDI lab water system produces Type II water from tap water directly. Resistivity of product water is above 5 MΩ·cm (25 ℃), which meets or exceeds Type Ⅱ water quality standard defined by ASTM, CAP, CLSI and ISO 3696/BS 3997, the European and U.S. Pharmacopoeia. The System is designed for users who need 400 to 1800 liters of Type II water per day.


Share
  • Type II Pure water can be used in many areas. Some of the typical applications are:
  • Preparation of bio-reagents
  • Preparation of culture media
  • Preparation of solution for chemical analysis such as HPLC, ICP etc.
  • For clinical analyzers
  • Feed water for ultrapure water systems
  • Medical device and equipment rinsing
  • For serum and blood fractions
  • For ophthalmics
Feed Water Requirements
 Tap water*  Municipal water, TDS < 1000 ppm (2000 μS/cm)
 Operating Temperature  5~45 ℃
 Feed Water Pressure  2 - 6 bar (30 - 90 psi)
 Product Water Flow Rate  Typically 30, 60, 125, 250 or 500 L/h (at 25 ℃)
 Product Water Quality
 Resistivity  > 5 MΩ·cm @ 25 ℃ (typically > 10 MΩ·cm @ 25 ℃)
 TOC  < 30 ppb
 Particle (> 0.2 μm)  < 1/ml (with a 0.2 μm final filter)
 Microorganisms  < 0.1 cfu/ml (with a 0.2 μm final filter)
 Pyrogens (endotoxins)  < 0.001 Eu/ml (with a terminal ultrafiltration cartridge)
 Dimensions
 Length x Depth x Height  40.0 cm X 49.0 cm X 110.0 cm
 Power  < 750 W

*If water hardness is greater than 120 ppm, RephiLe suggests using an additional water softening device to protect the RO membrane in your system.